|  | Fully automated, random access chemistry system with STAT capability | 
        
            |  | Spectrophotometry and potentiometry | 
        
            |  | Endpoint, rate, fixed point and indirect ISE | 
        
            |  | Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE | 
        
            |  | 60 photometric tests + 3 ISE, 120 pre-programmed onboard tests | 
        
            |  | 400 photometric tests/hour; maximum of 800 with ISE | 
        
            |  | 200 samples/hr. Maximum tests/hr: 600 if ISE only | 
        
            |  | Serum, plasma, urine, other | 
        
            |  | Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading | 
        
            |  | Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups | 
        
            |  | Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability. | 
        
            |  | 1–25 μl in 0.1 μl steps (1-25 μl for repeats) | 
        
            |  | 76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles. | 
        
            |  | Refrigerated (4–12°C) | 
        
            |  | R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment) | 
        
            |  | 90–350 μl | 
        
            |  | Permanent glass cuvettes | 
        
            |  | Up to 8 minutes, 37.5 seconds | 
        
            |  | 37°C, dry bath | 
        
            |  | Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible | 
        
            |  | 13 different wavelengths between 340–800 nm | 
        
            |  | Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode | 
        
            |  | Auto QC, cooled QC positions | 
        
            |  | 1. Auto rerun and manual rerun available2. Automatic sample increase, decrease or normal repeat
 3. Reflex testing
 | 
        
            |  | Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100 | 
        
            |  | Individual and profile test requisition via online, mouse, function keys and touch-screen | 
        
            |  | Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection. | 
        
            |  | RS232C. Full uni- and bidirectional communication possible. | 
        
            |  | Windows 7® | 
        
            |  | Sample number 100,000 samples. Reaction Data 200,000 tests. | 
        
            |  | ANL: 1450 mm x 1205 mm x 770 mm | 
        
            |  | 100–240v; 60 Hz/ < 3.5 kVA |